| Conductivity Probes The
conductivity measurement in electrolytes is made by means of a 2-electrodes-measuring
cell. A sine-shaped voltage with a frequency of about 2 kHz is put at the measuring
electrodes. The current flowing through the measuring object is converted into a voltage
which is phase-synchronously rectified, smoothed and then displayed related to 25°C.
The sensor is delivered completely aligned. Slight contamination can be removed by
means of a brush. After intensive cleaning of heavily soiled electrodes the electrode gaps
can slightly change and thus influence the measuring result. A re-alignment can be easily
carried out in two points at 0 mS/cm (dry) and 2.77 mS/cm or 147 mS/cm (ref solution) at
25°C by press of a key. 120 mm long shaft of PVC-C, 20 mm dia., electrodes of special
carbon, integrated electronics, NTC-temperature sensor for the conversion of the measuring
result related to 25°C.
FY A641-LF
-LF2 |
CONDUCTIVITY PROBE with autom. temperature compensation.
FY A641-LF: Laboratories 0 to 20 mS (linearized);
FY A641-LF2: pure water, 0 to 200 mS (linearized). |
 |
| Specifications for Conductivity
Probes |
| Operating temp. |
-5 to 70°C |
| Comp.-coefficient |
1.9 linear |
| Accuracy |
FY A641-LF: 0.01 - 5 mS/cm (± 1% of measured value + 0.05
mS);
5 - 20 mS/cm (± 2% of measured value + 0.05 mS);
FY A641-LF2: 5 - 200 mS/cm (± 1% of measured value + 0.5 mS) |
| Nominal temp. |
22°C ± 2 K |
| Res. cond./temp. |
FY A641-LF: 0.01 mS/cm, to choice 0.001 mS/cm/0.01°C;
FY A641-LF2: 0.1 mS/cm/0.01°C |
| Cell constant |
approx. 1 cm -1 |
| Min. immer. depth |
30 mm |
| Power supply |
approx. 8 V from ALMEMO® instrument/approx. 3 mA |
| Order No. |
Description |
| FY-A641-LF |
Laboratories, 0 to 20 mS (linearized) |
| FY-A641-LF2 |
Pure water, 0 to 200 µS (linearized) |
| ZA-9641-RL |
Reference solution (2.77 mS)
0.02 mol KCL 250 ml |
| ZA-9641-RL2 |
Reference solution (147 mS)
0.001 mol KCL 250 ml |