| Oxygen-sensor for O2 measurements
in gases Oxygen-sensor with screw-in thread for air-conditioning-plants, air-cleaner,
oxygen-rectifiers, greenhouses, oxygen-incubators, breathing apparatus, artificial
respiration instruments. The oxygen-sensor includes a lead-oxygen-cell, a lead-anode and a
gold-cathode under use of a special acid-electrolyte. The oxygen-molecules of the
gas-mixture exude through a non-porous teflon-membrane into the electro-chemical cell and
are there absorbed by the gold-electrode. The current flow between the electrodes is
proportional to the oxygen-concentration in the gas-mixture to be measured. The signals
are measured as voltage drop over the resistance and the NTC for the temperature
compensation. The modification of the output-voltage is proportional to the concentration
of the oxygen.
| Specifications for FY 9600-O2 |
| Mearusing range |
0 to 100% O2 |
| Output |
9 to 13 mV in air |
| Resolution |
0.01% O2 |
| Expected operating time |
2 years in 20.9% O2 |
| T99.5 response time |
< 40 seconds (equivalent to a change in concentration from 20.9% O2
to 0.1% O2 |
| Linearity |
linear 0 to 100% O2 |
| Baseline at 20°C |
< 20 µV |
| Temperature range |
-20 to 50°C |
| Pressure range |
atmospheric ±10% |
| Relative humidity range |
0 to 95% r.H. non condensing |
| Accuracy |
± 1% of O2 final value of measuring range |
| Cross sensitivity |
< 0.01% O2 in balance with other gases |
| Resistivity/cr |
no influence by CO2, Co, N2S, NOX, N2O
and H2 |
| Long term output drift |
< 2% signal/month (typically < 5 to 10 years) |
| Nominal conditions |
20°C, 50% r.H. 1013 mbar |
| Dimensions |
Ø 29,3 mm, length 43 mm, (Ø of absorbtion opening 10 mm) |
| Order No. |
Description |
| FY 9600-O2 |
O2 sensor for gases |
| ZA 9600-AK2 |
ALMEMO® adapter cable with 3.5 mm jack connector |