Carbon monoxide sensor for CO measurements in airfya600co.JPG (12767 bytes)

In garages for measuring and warnings tested acc. VDI 2053. For air quality control reg. max. working place concentration: e.g. in laboratories or motor testing stands. In homes and large shelters for surveillance of air outside or protection air.

The sensor includes a chemical measuring cell into which the sample diffuses. At the electrode the H+ -ions and the electrodes are used up in a cathode reaction. The power generated between anode and cathode is directly proportional to the CO concentration in the measured air.

 

Specifications for FY A600-CO
Mearusing range 0 to 300 ppm CO or 0 to 5% CO (50 000 ppm) other ranges upon request
Power supply from ALMEMO® instruments
Zero point error  < 10 ppm CO
Gauge reading bal. < 3 ppm CO
Accuracy ± 3% of final value
Zero point drift < 2% (1 year)
Repeatability  < 2% (1 year)
Set time T90 < 60 s
Transverse sensitivity < 2% by integrated sensor
Ambiant temp. -10 to 40°C temperature compensated
Relative humidity 0 to 99% non condensing
Durability approx. 2 years
Dimensions 80 mm Ø 80 mm high
Weight 600 g

 

Order No. Description
FY A600-CO Carbon monoxide sensor for CO measurements in air

 

{Back}