PC6
PRESSURE CALIBRATOR
The PC6 Pressure Calibrator combines the latest developments in pressure
instrumentation with advanced sensor and component technology to offer a versatile
calibration instrument.
It is designed for accurately testing and calibrating the following types of pressure
measuring instruments:
• Pressure Transducers
• Pressure Transmitters
• Pressure Switches
• Pressure Relief Valves
• Analogue Instrumentation (Pressure Gauges)
The PC6 is supplied with our hand held test systems:
TP1 Pneumatic pressure test system -1 to 25 bar
HTP1 Hydraulic pressure test system to 700 bar
The compact size of the PC6 does not compromise functionality. It can simultaneously show
pressure versus voltage, current, date & time, percentage of pressure, pressure switch
status
Features
• Low cost, small and portable
• Pressure measurement range -1 to 700 bar
• Voltage and current measurement
• Simultaneous display of pressure and selected function
• Timed / Event data storage with date & time stamping
• RS232 data communication link
• Pressure switch test
• Leak rate measurement 1 second to 999 days
• Maximum and minimum pressure measurement
• Display hold facility
• Supply output for transducer / transmitter
• Tare pressure offset values
• Pressure module temperature measurement
• Resolution variation facility
• External pressure module (EPM) interchangeability for required
pressure ranges
• Windows based software SiCal for remote control operation
and data transfer into many applications.
(Optional)
• CE marked
|
DC |
Input |
Range |
Resolution |
Accuracy |
Input resistance |
|
Voltage measurement |
mV |
0 to 100 mV |
1uV |
0.05% FS |
> 1 MW |
|
Current measurement |
V |
o to 5 V |
0.1 mV |
0.05% FS |
>1 MW |
The PC6 Indicator version does not include the Voltage/Current input/output or Switch functions.
| Specifications for PC6 | |
| Pressure units: | bar, Mpa, kPa, lb/in2, kg/cm2, atm, inH2O, mH2O, mmH2O,
inHg, mmHg, unit1, unit2 (user defined) |
| Overload Warning: | Flashing display and audible tone at 110% of FS |
| Temperature Display: | Temperature of pressure sensor module |
| Voltage Output: | Regulated DC supply 10V ± 0.1% max. current 30 mA Unregulated DC supply 18V ± 15% max. current 30 mA |
| Pressure Switch Test: | Status OPEN/CLOSED |
| Operating Temperature: | 0 to 40°C |
| Calibration Temperature: | 20°C ± 2°C |
| Calibrated Temp. Range: | 10 to 40°C |
| Storage Temperature: | -20 to 70°C |
| RS232 Parameters: | Baud rate 110, 300, 1200, 2400, 4800, 9600, stop bits 1 or 2 status ON or OFF |
| Zero Reset: | Manual-Keyboard operation |
| Data Capacity: | 176 records into a maximum of 20 files, 8k memory (additional 204 records, 8k memory option available) |
| Display: | Backlit LCD, 16 characters X 2 line alpha numeric |
| Pressure Media: | Water, Oil, Non-corrosive gases |
| Humidity: | 5 to 95% relative humidity non condensing |
| Resolution Select: | Increase or decrease by a factor of 10 |
| Power Supply: | 6V rechargeable battery / mains operation via charger (supplied) |
| Battery Life: | 10 hours fully charged |
| Low Battery: | Continuous check, audio and visual warning |
| Dimensions: | 92 x 110 x 59 mm |
| Weight: | 1.2 kg |
| Electrical Connections: | Ground:
Standard 2 mm socket Voltage / Current Input: Standard 2 mm socket Voltage Output: Standard 2 mm socket Pressure Switch Input: Standard 2 mm socket Battery Charger: 2-pole miniature round connector RS232: 8-pole miniature round connector EPM: 14-pole miniature round connector |
| Pressure Connection: | 1/4 BSP male quick-fit
(Pneumatic) 3/8 BSP femal quick-fit (Hydraulic) |
| Software (optional): | SICal will allow users to control the
PC6 Calibrator remotely through a virtual instrument panel on the computer screen. The
calibrator data can logged, viewed and printed in graphical format or on a calibration
certificate. System Requirements: 486 with maths coprocessor. Windows 3.1 or higher. |
| Order No. Description | |
| TP1-HI | (DIGITAL GAUGE .05% KIT) |
| HTP1-HI | (DIGITAL GAUGE .05% KIT) |
| TP1-PC6 | (DIGITAL CAL .05% KIT) |
| HTP1-PC6 | (DIGITAL CAL .05% KIT) |
|
(For Indicators, the suffix C is replaced with an I)
HTP1-KIT include: Pump, digital gauge 0.05% accuracy, 7 piece adaptor set and seals, hydraulic hose assembler,1/4 BSP quick fit connection |
|
|
TP1-KIT include: Pneumatic pump, 0.05% PC6I datalogging indicator, hose plus 1/4 BSP swivel adaptor, 7 piece adaptor set, plastic case and foams |
|
{back}